July 2018

World Class 3000 Intelligent Field Transmitter Setup

The DR probe operates with a 115 VAC heater. Ensure that the voltage selection jumpers in
the IFT or HPS, if used, are set properly. Refer to Figure 2-14 for IFT jumper selection, and
figure 2-15 for HPS jumper selection. For additional setup and configuration information,
refer to IB-106-300NF or IB-106-300NH.
Figure 2-14. IFT 3000 Power Supply Board Jumper Configuration
Line Voltage
Selection
100 VAC
120 VAC
220 VAC
240 VAC
CAUTION
If incorrect heater voltage is selected, damage to the Oxymitter DR may occur. refer to IB-106-300NH or
IB-106-300NF for additional configuration information. Always update the relevant labeling to reflect the set
voltage.
Figure 2-15. Heater Power Supply (HPS 3000) Jumper Configuration
Line Voltage
Selection
100/120 VAC
1
220/240 VAC
Voltage Selection
World Class Probe (44V)
218 Probe (115V)
Direct Replacement World
Class or DR Oxymitter
100 VAC operastion requires transformer part number 1M02961G02.
1
Installation
Jumper
(Install)
JM3, JM7, JM2
JM8, JM7, JM1
JM6, JM5, JM2
JM6, JM5, JM1
Jumper
(Install)
JM4, JM1
JM5
Probe Heater
Jumper
(Install)
Jumper
Always disconnect line voltage
configuration
from intelligent field transmitter
Probe Heater
Voltage Selection
World Class Probe (44V)
218 Probe (115V)
World Class "Direct Replace-
ment" Probe (115V)
or Oxymitter Direct
Replacement Probe (115V)
Always disconnect line voltage
from heater power supply and
analog electronics (if used) before
Jumper
Configuration
Heater Power
Remote
On
Electronics Selection
*Analog (Existing)
JM7
Digital
(Next Generation)
JM8
JM8
Rosemount Oxymitter 4000
before changing jumpers.
Jumper
(Install)
JM10
JM9
JM9
changing jumpers.
Jumper
Remove JM2
*Install JM2
Jumper
Install JM3, JM6
Remove JM3, JM6
36210015
36210014