Figure 8-8. Fault 6, High High Heater Temp
LOI
When Fault 6 is detected, the LOI displays the "Very Hi O
1.
2.
3.
4.
Troubleshooting
The very high O
temperature alarm will activate when the thermocouple
2
produces a voltage of 37.1 mV [900°C (1652°F)].
The triac and the temperature control may be at fault.
Remove power. Allow the Oxymitter 4000 to cool for five minutes. Restore power.
If the condition repeats, replace the electronic assembly per "Electronic Assembly
Replacement" in Section 9: Maintenance and Service.
Restore power.
4. If the condition repeats, repla