Figure 9-10. Ceramic Diffuser Element Replacement
8. Apply a light coating of anti-seize compound to both sides of the new
corrugated seal.
9. Assemble the cell and flange assembly and corrugated seal to the
probe tube. Make sure the calibration tube lines up with the calibration
gas passage in each component. Apply a small amount of anti-seize
compound to the screw threads and use the screws to secure the
assembly. Torque to 4 Nám (35 in-lbs).
10. Apply anti-seize compound to the probe threads, flame arrestor hub,
and setscrews. Reinstall the flame arrestor on the probe. Using pin
spanner wrenches, torque to 14 Nám (10 ft-lbs). Secure the flame
arrestor with the locking setscrew. Torque to 2.8 Nám (25 in-lbs). If
applicable, reinstall the vee deflector, orienting apex toward gas flow.
Secure with the setscrew and anti-seize compound. Torque to 2.8 Nám
(25 in-lbs).
11. On systems equipped with an abrasive shield, install the dust seal
gaskets, with joints 180¡ apart.
12. If previously removed, install the entire electronics per "Replacement of
Entire Electronics (with Housing)", steps 4 and 5.
13. Follow the instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 4000 into the stack or duct. If there
is an abrasive shield in the stack, make sure the dust seal gaskets are in
place as they enter the 15¡ reducing cone.
152
M5-0.8 X 5 mm
Locking Set Screw
(Use 2.5 mm Hex Key)
Threaded
Probe Flange
Flame
Arrester
Hub
Cement
Port
Cement
8.
Apply a light coating of anti-seize compound to both sides of the new
corrugated seal.
9.
Assemble the cell and flange assembly and corrugated seal to the probe
tube. Make sure the calibration tube lines up with the calibration gas
passage in each component. Apply a small amount of anti-seize compound
to the screw threads and use the screws to secure the assembly. Torque to 4
N·m (35 in-lbs).
10.
Apply anti-seize compound to the probe threads, flame arrestor hub, and
setscrews. Reinstall the flame arrestor on the probe. Using pin spanner
wrenches, torque to 14 N·m (10 ft-lbs). Secure the flame arrestor with the
locking setscrew. Torque to 2.8 N·m (25 in-lbs). If applicable, reinstall the
vee deflector, orienting apex toward gas flow. Secure with the setscrew and
anti-seize compound. Torque to 2.8 N·m (25 in-lbs).
11.
On systems equipped with an abrasive shield, install the dust seal gaskets,
with joints 180°apart.
12.
If previously removed, install the entire electronics per "Replacement of
Entire Electronics (with Housing)", steps 4 and 5.
13.
Follow the instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 4000 into the stack or duct. If there
is an abrasive shield in the stack, make sure the dust seal gaskets are in place
as they enter the 15° reducing cone.
Instruction Manual
IM-106-340C, Rev. 6.2
Spanner
Wrench
Ceramic
Diffusion
Element
M6-1 X 6 mm
Setscrew
(Use 3 mm
Hex Key)
Fillet
Vee
Deflector
Apr 2015
Maintenance and Service
9-20