Calibration Gas; Imps 4000 Connections; Sps 4001B Connections - Emerson Rosemount Oxymitter 4000 Reference Manual

Hazardous area oxygen transmitter
Hide thumbs Also See for Rosemount Oxymitter 4000:
Table of Contents

Advertisement

July 2018
Figure 2-12. Hazardous Area Oxymitter 4000 Gas Connections

Calibration Gas

Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low
(zero) be between 0.4% and 2.0% O
of more than 40 parts per million. Failure to use proper gases will result in erroneous
readings.
Two calibration gas concentrations are used with the Hazardous Area Oxymitter 4000,
Low Gas - 0.4% O
Oxymitter 4000 connections.
If using an IMPS 4000 or an SPS 4001B, install it in a non-hazardous, explosive-free
environment.

IMPS 4000 connections

Ensure the IMPS 4000 is installed in a safe (non-hazardous, explosive-free) area and verify
the wiring and pneumatic connections per the IMPS 4000 Intelligent Multi-probe Test
Gas Sequencer Instruction Manual.

SPS 4001B connections

Ensure the SPS 4001B is installed in a safe (non-hazardous, explosive-free) area and verify
the wiring and pneumatic connections per the SPS 4001B Single Probe Autocalibration
Sequencer Instruction Manual.
Installation
Hazardous Area Oxymitter 4000
Calibration Gas
WARNING
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero)
be between 0.4% and 2.0% O
than 40 parts per million. Failure to use proper gases will result in erroneous readings.
Two calibration gas concentrations are used with the Hazardous Area
Oxymitter 4000, Low Gas - 0.4% O
for the Hazardous Area Oxymitter 4000 connections.
and High Gas - 8% O
2
WARNING
If using an IMPS 4000 or an SPS 4001B, install it in a non-hazardous, explosive-free
environment.
Ensure the IMPS 4000 is installed in a safe (non-hazardous, explosive-free)
area and verify the wiring and pneumatic connections per the IMPS 4000
Intelligent Multi-probe Test Gas Sequencer Instruction Manual.
Ensure the SPS 4001B is installed in a safe (non-hazardous, explosive-free)
area and verify the wiring and pneumatic connections per the SPS 4001B
Single Probe Autocalibration Sequencer Instruction Manual.
NOTE
Upon completing installation, make sure that the Hazardous Area Oxymitter
4000 is turned on and operating prior to firing up the combustion process.
NOTE
Damage can result from having a cold Hazardous Area Oxymitter 4000
exposed to the process gases. During outages, and if possible, leave all
Upon completing installation, make sure that the Hazardous Area Oxymitter 4000 is
Hazardous Area Oxymitter 4000 units running to prevent condensation and
turned on and operating prior to firing up the combustion process. Damage can
premature aging from thermal cycling.
result from having a cold Hazardous Area Oxymitter 4000 exposed to the process
gases. During outages, and if possible, leave all Hazardous Area Oxymitter 4000
units running to prevent condensation and premature aging from thermal cycling.
WARNING
If the ducts will be washed down during outage, MAKE SURE to power down the Hazardous
If the ducts will be washed down during outage, MAKE SURE to power down the
Area Oxymitter 4000 units and remove them from the wash area.
Hazardous Area Oxymitter 4000 units and remove them from the wash area.
R
SMART
800-433-6076
FAMILY
HART T
M
OXYMITTER 4000
SERIAL NO.
TAG NO.
VOLTS:
85-264 VAC W 4 8 A -
500 VA
6 T 2 T H S z : OUTPUT: 4-20
5 Amps
mALINE FUSE:
Reference Air
. Do not use gases with hydrocarbon concentrations of more
2
. Do not use gases with hydrocarbon concentrations
2
and High Gas - 8% O
2
. See Figure 2-12 for the Hazardous Area
2
Rosemount Oxymitter 4000
. See Figure 2-12
2
55