Figure 2-8. Remote Electronics Mounting
44

Remote Electronics Installation

For a Hazardous Area Oxymitter 4000 with the remote electronics option,
For a Hazardous Area Oxymitter 4000 with the remote electronics option, install the
install the probe according to the instructions in ÒProbe InstallationÓ on
probe according to the instructions in "Probe Installation" on page 2-2. Install the
page 2-2. Install the remote electronics unit on a wall, stand pipe, or similar
remote electronics unit on a wall, stand pipe, or similar structure (Figure 2-2 and Figure
structure (Figure 2-2 and Figure 2-8).
2-8).

Electrical Installation with Integral Electronics

All wiring must conform to local and national codes.
All wiring must conform to local and national codes.
WARNING
Disconnect and lock out power before connecting the unit to the power supply.
Disconnect and lock out power before connecting the unit to the power supply.
WARNING
Install all protective equipment covers and safety ground leads after installation. Failure
to install covers and ground leads could result in serious injury or death.
WARNING
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 61010 (EC requirement), and ensure safe
operation of this equipment, connection to the main electrical power supply must be
made through a circuit breaker (min 10 A) which will disconnect all current-carrying
conductors during a fault situation. This circuit breaker should also include a
mechanically operated isolating switch. If not, then another external means of
disconnecting the supply from the equipment should be located close by. Circuit
To meet the Safety Requirements of IEC 61010 (EC requirement), and ensure safe
operation of this equipment, connection to the main electrical power supply must be made
breakers or switches must comply with a recognized standard such as IEC 60947.
through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors
during a fault situation. This circuit breaker should also include a mechanically operated
isolating switch. If not, then another external means of disconnecting the supply from the
WARNING
equipment should be located close by. Circuit breakers or switches must comply with a
recognized standard such as IEC 60947.
The probe and probe abrasive shield are heavy. Use proper lifting and carrying
procedures to avoid personnel injury.
The probe and probe abrasive shield are heavy. Use proper lifting and carrying procedures
to avoid personnel injury.
July 2018
Instruction Manual
IM-106-340C, Rev. 6.2
Apr 2015
Installation
2-10