EDAN F15 Series User Manual page 81

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
FT20 MALF
Current
network
environment
poor(FT20)
FT20
Wireless
Signal Weak
ECG LEADS OFF
ECG SINGNAL
EXCEEDS LIMIT
Bad
contact
between
the
electrode
sleeve
and the skin
ECG Signal Weak
HR
EXCEEDS
MEASUREMENT
RANGE
NIBP
EQUIP
MALF
NIBP
SYSTEM
FAILURE
NIBP
Airway
Pressure
Abnormality
NIBP
CUFF
LOOSE or OFF
NIBP
OVER
PRESSURE
014_14.1_F15_Series_User_Manual
FT20 is disconnected with the
monitor
The current network environment
is poor and FT20 module has
problem to connect with the
monitor
The wireless signal of FT20
module is too weak for the system
to analyze.
The ECG cable is not well
connected to the transducer, or
the leads of the ECG cable are not
well attached to the patient.
ECG
signal
exceeds
measurement limits.
Bad
contact
between
electrode
sleeve
transducer and the skin.
The ECG signal is too weak that
the system can't calculate HR
correctly.
The
heart
rate
exceeds
measurement limits.
The
NIBP
board
communicate with the system
successfully.
The NIBP module defective.
NIBP cannot be deflated, and the
system pressure has remained
unchanged
The cuff is loose or not connected. Wrap the cuff properly.
The pressure has exceeded the
specified upper safety limit.
Restart the monitor and try again, contact
the manufacturer if the connection still
fails.
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Check whether the ambulating mother
has reached the limit of the active signal
area-of-reach and whether the network
communication is fine.
Check the connection between the ECG
cable and the monitor or the TOCO-E
transducer. Check also the attachment of
the ECG leads to the patient.
the
Check the connection of the leads and the
patient's condition.
the
Adjust the transducer and make sure
of
TOCO
good contact between the electrodes and
the skin.
Check the attachment of the ECG leads to
the patient.
Check the connection of the ECG
the
leads/SpO
condition.
cannot
Restart the monitor and try again, contact
the manufacturer if the connection still
fails.
Restart the monitor and try again, contact
the manufacturer if the connection still
fails.
Measure again, if the fault persists, stop
using the monitor for NIBP measurement,
and contact the manufacturer for repair
Measure again, if failure persists, stop
using the monitor for NIBP measuring and
contact the manufacturer for service.
- 65 -
transducer and the patient's
2
Alarms
81/202