Monitoring Uterine Activity Internally; Parts Required; Directions For Use Of Iupc - EDAN F15 Series User Manual

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
1
Switch on the monitor.
2
Connect the TOCO transducer to the fetal monitor.
3
Gently press the center of the transducer.
4
Check that the value on the display shows this change in pressure.
If a TOCO transducer fails the test, repeat this test with another transducer. If the second one passes
the test, defect of the first transducer is confirmed. Replace it with a good one. If the second
transducer fails the test as well, contact the manufacturer for service.

7.7 Monitoring Uterine Activity Internally

7.7.1 Parts Required

a)
Disposable intrauterine pressure catheter ACCU-TRACE
b) DECG-IUP Integrated Cable
c)
Wired/wireless DECG Fetal&Maternal Module
The following illustration shows how these parts should be connected:
(a)

7.7.2 Directions for Use of IUPC

Preparation
1)
Gather supplies: ACCU-TRACE IUPC, DECG-IUP integrated cable, and amnioinfusion
supplies if needed.
2)
Open the sterile ACCU-TRACE IUPC package.
Insertion
NOTE:
014_14.1_F15_Series_User_Manual
Figure 7-7 Testing a TOCO Transducer
(b)
Figure 7-8 Connection for IUP Monitoring
- 90 -
IUPC ("IUPC" for short)
(c)
Fetal Monitoring