Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
Note 1:
1. The wireless signal shall cover all marked labour rooms and delivery rooms.
2. The wall of inpatient ward is thick brick wall, the roof of corridor and rooms is the movable ceiling,
with material gypsum, height 3 m from the ceiling to the ground.
3. The wall of delivery room is thick brick wall, with the door made of metal.
4. The toilet position is as shown in the figure.
出口
1000
Fetal Monitor
Fetal Monitor
Fetal Monitor
Labour
Fetal Monitor
Fetal Monitor
Room
Fetal Monitor
Fetal Monitor
Bearing wall
Brick wall

A9.5 Network Design and Implementation

A9.5.1 Networking Scheme

F15 mainly involves networking between the bedside monitor and the central station; F15 Air not only
involves networking between the bedside monitor and the central station, but also networking between
the wireless transducer and monitor.
Wireless transducers and monitors have the following requirements for wireless coverage:
NO.
Item
1
Equipment compatibility
2
802.11 protocol
Spatial signal strength
3
4
Co-channel interference
Adjacent
5
interference
014_14.1_F15_Series_User_Manual
Network Plan
Network of XX Hospital
Dispensary
Office
Nurse Station
Central Station
Fetal Monitor
Toilet
Fetal Monitor
VIP
Labour
Room
Fetal Monitor
Cavity wall
Monitor Requirements
It is recommended to use the
model recommended by EDAN
802.11 a/b/g/n requirements are
met
≥-65dBm
The signal coverage requirement
of the AP connected to the
monitor is the RSSI value felt by
the monitor.
≤ -20dB
Measurement within the monitor's
working channel
≤15dB
Measurement
frequency
monitor's working channel
Recommended
improve the SNR of Wi-Fi
- 175 -
WI-FI Networking and Instructions
Fetal
Fetal
Monitor
Monitor
Delivery
Delivery
Room
Room
Labour
Labour
Room
Room2
Toilet
Fetal Monitor
Toilet
Fetal Monitor
Fetal Monitor
60 m
around
the
to
meet,
can
Rest
Fetal
Room
Monitor
Delivery
Room
Labour Room
Labour Room
Toilet
Fetal Monitor
Fetal Monitor
Description
Recommended
to
meet
Required to meet.
Required to meet.
Required to meet.
Required to meet.