Cleaning The Ecg Cable Assembly; Cleaning The Toco Transducer And Spo2 Sensor; Cleaning The Temp Sensor; Disinfecting - EDAN F15 Series User Manual

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
3.
Adjust the bladder until it is in position.
1 Do not squeeze the rubber tube on the cuff.
2 Do not dry-clean the cuff.
3 Only clean the outer surface of the connectors, make sure no liquid goes into the
connector.
4 When the reusable cuff is not connected with the monitor, or being cleaned, always place
the cover on the rubber tube to avoid liquid permeation.

11.2.2.3 Cleaning the ECG Cable Assembly

1.
Wipe the cable assembly with a soft cloth dampened with the cleaning solution until no visible
contaminants remain.
2.
After cleaning, wipe off the cleaning solution with a fresh cloth or towel dampened with tap water
until no visible cleaning agent remains.
3.
Wipe off residual moisture with a dry cloth.
4.
Leave the cable assembly to air dry.

11.2.2.4 Cleaning the TOCO Transducer and SpO2 Sensor

1.
Wipe the surfaces of the sensor and cable using a soft cloth dampened with the cleaning solution
until no visible contaminants remain.
2.
Wipe the patient contact area of the sensor with the cotton swab dampened with the cleaning
solution until no visible contaminants remain.
3.
After cleaning, wipe off the cleaning solution with a fresh cloth or towel dampened with tap water
until no visible cleaning agent remains.
4.
Wipe off residual moisture with a dry cloth.
5.
Leave the sensor to air dry.

11.2.2.5 Cleaning the TEMP Sensor

1.
Wipe the patient contact area with a soft cloth dampened with the cleaning solution until no
visible contaminants remain.
2.
After cleaning, wipe off the cleaning solution with a fresh cloth or towel dampened with tap water
until no visible cleaning agent remains.
3.
Wipe off residual moisture with a dry cloth.
4.
Leave the sensor/probe to air dry.

11.3 Disinfecting

Clean the monitor and reusable accessories before they are disinfected. The validated disinfectants
for cleaning the monitor and reusable accessories are:
Ethanol (75%)
014_14.1_F15_Series_User_Manual
CAUTION
- 126 -
Maintenance and Cleaning
142/202

Advertisement

Table of Contents
loading

Table of Contents