File Management (Optional) - EDAN F15 Series User Manual

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
3
Select ID, input the patient ID with the soft keyboard and select Enter.
Select Name, input the patient's name with the soft keyboard and select Enter.
4
NOTE:
You can input only a part of the patient ID or name. However, the more information you
input, the more accurate result you will get.
5
Select Search. The files with the matched information are listed in the window.
6
Select the required item, and this file is loaded to the main interface immediately. You can review
the traces backward or forward.

3.6.3.3 * File Management (Optional)

The USB facility of the monitor allows you to export the auto-saved files into a USB disk, and then you
can save the files in a PC or open them in a data managing system.
Once the monitor is configured with the relevant hardware, the USB feature can be enabled or
disabled by the service personnel of the manufacturer.
To Export files,
1
Make sure the USB feature is enabled. Stop printing and disconnect the network.
2
Plug the USB disk into the USB socket on rear panel of the monitor (figure 3-5). A message
"Ready to use USB disk" in the message area indicates the proper insertion of the disk.
3
Select the file managing key
Management interface, which records a list of up to 300 most recent monitoring records (patients'
ID, name and date) and a few operation items.
014_14.1_F15_Series_User_Manual
Figure 3-37 Patient Searching
- 44 -
Basic Operation
on the main interface to open the File
60/202

Advertisement

Table of Contents
loading

Table of Contents