A4.6 Transducer Acoustic Output Parameters List; A4.6.1 Test Of Wired Transducer(F15) - EDAN F15 Series User Manual

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual

A4.6 Transducer Acoustic Output Parameters List

A4.6.1 Test of Wired Transducer(F15)

(IEC60601-2-37:2007+AMD1:2015, table 201.103)
Transducer Model: US-TF
Nominal Frequency: 1.0MHz
Operating Mode: PW Mode
Index label
Maximum index value
Index component value
Acoustic
p
at z
r.α
MI
Parameters
(MPa)
P
P
1x1
z
s
z
b
z
MI
z
PII.α
f
awf
Other
prr
Information
srr
n
pps
I
pa.α
2
(W/cm
)
I
at z
spta.α
z
(mW/cm
SII.α
I
at z
spta
(mW/cm
p
at
r.
(MPa)
Operating
Focus(mm)
control
Depth(mm)
conditions
Frequency(MHz)
NOTE: N/A indicates that there is no corresponding intended use or no data reported.
Acoustic Output Reporting Table for Track 1(Non-autoscanning Mode)
Transducer Model: US-TF
Nominal Frequency: 1.0MHz
Operating Mode: PW Mode
Acoustic Output
Global Maximum Value
p
P
Associated
f
c
Acoustic
Z
Parameter
Beam
dimensions
PD
014_14.1_F15_Series_User_Manual
Acoustic output reporting table
MI
0.034
0.034
(mW)
(mW)
(cm)
(cm)
(cm)
0.50
(cm)
0.50
(MHz)
1.00
(Hz)
2000.00
(Hz)
N/A
N/A
at
z
0.056
PII.α
9.97
or
PII.α
2
)
or z
10.32
PII
SII
2
)
z
0.035
PII
Fixed
Fixed
1.00
(MPa)
r.3
(mW)
o
(MHz)
(cm)
sp
x
-6
y
-6
(usec)
TIS
At
Below
surface
surface
0.0034
N/A
0.0034
16.43
N/A
5.30
1.00
N/A
Fixed
N/A
Fixed
N/A
1.00
MI
0.034
0.034
1.00
0.50
(cm)
(cm)
90.97
- 150 -
Ultrasound Intensity and Safety
TIB
At
Below
surface
surface
0.12
N/A
0.12
16.43
N/A
5.30
1.00
N/A
Fixed
N/A
Fixed
N/A
1.00
I
SPTA.3
2
(mW/cm
)
9.97
16.43
1.00
0.50
0.42
0.40
TIC
N/A
N/A
N/A
N/A
N/A
N/A
I
SPPA.3
2
(W/cm
)
0.056
16.43
1.00
0.50
0.42
0.40
90.97