EDAN F15 Series User Manual page 76

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
Low Priority
Check Paper
US1 UNPLUGGED or
US2 UNPLUGGED or
US3 UNPLUGGED
US1 Signal Weak or
US2 Signal Weak or
US3 Signal Weak
FETUS 302 MODULE
MALF or Extend 302
MODULE
MALF
FETUS 417 MODULE
MALF
TOCO UNPLUGGED
IUP LEADS OFF
DECG LEADS OFF
DECG UNPLUGGED
FECG RLD Off
FECG
LEADS
OFF(TOP/BOTTOM/C
4/F/R/N/L)
DECG Signal Weak
US1 Wireless Signal
Weak
US2 Wireless Signal
Weak
US3 Wireless Signal
Weak
014_14.1_F15_Series_User_Manual
transducer.
There is no paper in the paper
drawer or the drawer is open.
US
transducer
transducer 2 or US transducer
3 is not well connected.
FHR1 or FHR2 or FHR3 signal
is too weak for the system to
analyze.
The fetus 302 module or
extend 302 module or fetus
or
407 cannot communicate with
the system successfully.
TOCO transducer is not well
connected.
DECG-IUP integrated cable is
not well connected.
The spiral electrode or DECG
cable is not well connected.
The
DECG
Fetal&Maternal
Module is not well connected.
FECG electrodes (C1//N/F) are
not well connected.
FECG
(top/bottom/C4/F/R/N/L)
not well connected.
DECG signal is too weak for
the system to analyze.
Wireless FHR1 signal is too
weak
for
the
analyze.
Wireless FHR2 signal is too
weak
for
the
analyze.
Wireless FHR3 signal is too
weak
for
the
analyze.
Load paper and/ or close the drawer.
1
or
US
Check the connection of the transducer.
Check if the US transducer is aimed at
the fetal heart; check if the alarm limits
are
suitable;
condition.
Restart the monitor and try again,
contact
connection still fails.
Check the connection of both TOCO
transducer and US transducer.
Check the connection of the DECG-IUP
integrated cable.
Check the connection of the spiral
electrode or DECG cable.
Check
Fetal&Maternal Module.
Check the connection of the FECG
electrodes.
leads
Check the connection of the FECG
are
leads (top/bottom).
Check if the spiral electrode is well
attached to the fetus; check the patient's
condition.
Check whether the ambulating mother
has reached the limit of the active signal
system
to
area-of-reach and whether the network
communication is fine.
Check whether the ambulating mother
has reached the limit of the active signal
system
to
area-of-reach and whether the network
communication is fine.
Check whether the ambulating mother
has reached the limit of the active signal
system
to
area-of-reach and whether the network
communication is fine.
- 60 -
check
the
the
manufacturer
the
connection
of
Alarms
patient's
if
the
DECG
76/202