EDAN F15 Series User Manual page 79

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
Patient Alarm Messages
Alarm Message
High Priority
***ASYSTOLE
Medium Priority
**HR HIGH or
**HR xxx > yyy
**HR LOW or
**HR xxx < yyy
** SpO
HIGH or
2
** SpO
xxx > yyy
2
** SpO
LOW or
2
** SpO
xxx < yyy
2
**SYS HIGH or
**SYS xxx > yyy
**SYS LOW or
**SYS xxx < yyy
**DIA HIGH or
**DIA xxx > yyy
**DIA LOW or
**DIA xxx < yyy
**MAP HIGH or
**MAP xxx > yyy
**MAPLOW or
**MAP xxx < yyy
**TEMP HIGH or
**TEMP xxx > yyy
**TEMP LOW or
**TEMP xxx < yyy
Technical Alarm Messages
Alarm Message
High Priority
***
CPU
Temp
high
014_14.1_F15_Series_User_Manual
Cause
No QRS wave is detected by
TOCO&MECG transducer in 4
seconds or no QRS wave is
detected
by
Fetal&Maternal
Module
seconds
Maternal HR result (xxx) is higher
than the upper limit (yyy).
Maternal HR result (xxx) is lower
than the lower limit (yyy).
SpO
result (xxx) is higher than
2
the upper limit (yyy).
SpO
result (xxx) is lower than the
2
lower limit (yyy).
SYS result (xxx) is higher than the
upper limit (yyy).
SYS result (xxx) is lower than the
lower limit (yyy).
DIA result (xxx) is higher than the
upper limit (yyy).
DIA result (xxx) is lower than the
lower limit (yyy).
MAP result (xxx) is higher than the
upper limit (yyy).
MAP result (xxx) is lower than the
lower limit (yyy).
TEMP result (xxx) is higher than
the upper limit (yyy).
TEMP result (xxx) is lower than
the lower limit (yyy).
Cause
CPU temperature exceeds the
safe limits
Countermeasure
Check the patient's condition and take
FECG
necessary measures.
in
6
Check if the alarm limits are suitable;
check the patient's condition.
Countermeasure
Check whether the fan of the monitor
works normally, if no, please contact the
manufacturer for service
- 63 -
Alarms
79/202