Data Saving; Searching For A File - EDAN F15 Series User Manual

Fetal & maternal monitor
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Traditional 510(k) of Fetal & Maternal Monitor
F15 Series Fetal & Maternal Monitor User Manual
1
Figure
1 Searching Key
3 Alarm Reviewing Key

3.6.3.1 Data Saving

When the START key is pressed, the monitor saves data of the previous ID in a file, and then clears it
from the main interface. The main interface only displays the new patient's data. During monitoring,
the data is saved every 10 minutes. All data of the same patient is saved in a file (the maximum
duration is 24 hours, the rest data is saved in another file.)
The files are stored in the monitor. When the data amount reaches the maximum capacity (500 files or
5000 hours), the monitor deletes the oldest file(s) automatically.
When the network version is ETHERNET 1.4 and above, if the central station disconnects with the
fetal monitor in one minute, the monitoring data will be re-uploaded to the central station after
re-connection, if the disconnection is over one minute, then the monitoring data can't be uploaded to
the central station.

3.6.3.2 *Searching for a File

When the USB facility is disabled, the searching key
used to search for a patient's data file saved in the monitor.
To search for a patient,
Select the searching key to open the file list. It contains six sets of most recent patients' ID, name
1
and start time of monitoring. Select the required item, this file is loaded to the main interface
immediately.
If the required file is not in this list,
2
Select MORE to open the Patient Searching window.
014_14.1_F15_Series_User_Manual