Appendix; Error Messages - Rice Lake LaserLight3 Series Technical Manual

Remote display
Hide thumbs Also See for LaserLight3 Series:
Table of Contents

Advertisement

LaserLight3 Remote Display
9.0

Appendix

9.1

Error Messages

The LaserLight3 provides a number of front panel error messages to assist in problem diagnosis.
messages and their meanings.
Error Message
-------
v v v v v v v
Battery low
Waiting for 30 seconds of standstill
Not allowed in motion
Tare in motion not allowed
Tare not allowed
Negative tare not allowed
Keyed tare not allowed
Tare exceeding capacity not allowed Displays when attempting to perform a tare larger than capacity, if it is not allowed
Multiple tares not allowed
Must be in first interval to tare
Keyed tare value must be in first
interval
Initial zero failed
Please wait
Alarm
NODATA
88
Visit our website
Overflow error – Weight value too large to be displayed
Gross > overload limit – Gross value exceeds overload limit; Check configuration or signal input level; Overload can
be caused by input signal > 45 mV or common mode voltage > 950 mV
Gross < underload limit – Gross value exceeds underload limit
Displays every 30-seconds when CR2032 battery is 2.7V or less
Displays when power up delay is active
Displays when attempting to perform a function while weight is in motion, if in motion is not allowed for that function
Displays when attempting to perform a tare while weight is in motion, if in-motion tares are not allowed
Displays when attempting to perform a tare, if a tare is not allowed
Displays when attempting to perform a negative tare, if a negative tare is not allowed
Displays when attempting to perform a keyed tare, if a keyed tare is not allowed
Displays when attempting to perform a tare if a tare is already in the system, if tare is not configured to
replace or remove
Interval is currently within the second or third interval; Perform tare while in the first interval only
Keyed tare value is within the second or third interval; Keyed tare value must be within the first interval only
Displays when an attempt to perform an initial zero fails, only possible at startup
Displays when calibrating
Displays when a configured setpoint's ALARM action occurs
Display when a data stream is not currently received by the LaserLight3. Ensure Indicator is transmitting a data
stream to LaserLight3, LaserLight3 is set to RmtIn (Remote In), and wiring is correct/undamaged.
Table 9-1. LaserLight3 Error Messages
www.RiceLake.com
Table 9-1
Description
lists these