Table 3-1. Parameters Calculated by the Analyzer
Parameter
Inspiratory Time
Expiratory Time
Inspiratory Hold Time
Expiratory Hold Time
Inspiratory to Expiratory
Time
Inspiratory Hold
Pressure
Positive End-Expiratory
Pressure
Mean Airway Pressure
Peak Inspiratory
Pressure
Tidal Volume
Minute Volume
Peak Inspiratory Flow
Rate
Peak Expiratory Flow
Rate
Compliance
Abbreviation
In Time
Time of the inspiratory period including the
inspiratory hold time
Ex Time
Time of the expiratory period including the
expiratory hold time
In Hold
Time of the End Inspiratory State
Ex Hold
Time of the End Expiratory State
Ratio I:E
Ratio of the inspiratory time to the Expiratory
Time
IPP
Average pressure during the End Inspiratory
State
PEEP
Average pressure during the End Expiratory
State
MAP
Mean pressure over the entire breath
PIP
Maximum pressure during the Inspiratory Time
Tidal Vol
Total volume accumulated over the Inspiratory
Time
Minute Vol
Expiratory Tidal Volume times the respiratory
rate in breaths per minute; reported as an eight-
breath average
PIF
Maximum of the absolute value of the Flow Rate
measured during the Inspiratory Time
PEF
Maximum of the absolute value of the Flow Rate
measured during the Expiratory Time
N/A
Expiratory Tidal Volume divided by the difference
of the Inspiratory Hold Pressure and PEEP