1.
Select Calibration on the right side of the Operation screen. The Calibration Setup window
appears.
2.
On the Setup tab, set up the blend concentrations.
Figure 7-1: Set up gas concentrations
3.
Click Send.
4.
From the Operation screen, select Cal.
Figure 7-2: Run calibration
5.
When the current cycle completes, the device should begin a calibration on the designated
calibration stream (default is stream 4).