Figure 6-29: Run the calibration stream (Stream 4)
Figure 6-30: Calibration stream analysis running
2.
Allow the stream to process for two or three cycles (approximately 10 to 15 minutes). During
the final cycle, change the next mode to Hold. When the device completes the current cycle, it
will enter hold mode.
3.
Select Next to verify analysis results.
4.
At the Results tab, compare normalized percent for each component to the component and
percent listed on the calibration blend cylinder. Component percentages should be the same.
21 01 51 0MN A H | NG C 82 0 0 A N D P GC 10 00 US ER MA NUA L | 6 5