Figure 4-15: Sample conditioning module line connections
IMPORTANT NOTE: Tube, ferrule and nut should always enter the connection at a right angle.
5.
If necessary, install the reducer into the sample probe output fitting.
6.
Install the ferrule and nut onto one end of the sample tubing.
7.
Insert the tubing with the ferrule into the reducer/sample probe output fitting. Move the nut
down onto the ferrule, screw onto fitting and tighten.
8.
Install the ferrule and nut onto the other end of the sample tubing.
9.
Insert the ferrule into the sample conditioning module input fitting. Move the nut down onto the
ferrule, screw onto fitting and tighten.
10. Locate the sample output fitting on the sample conditioning module.
11. Locate the sample input on the feed-through assembly and remove the sealing screw.
NOTICE – Equipment damage. Leave the sealing screw in any unused ports. If unused
stream ports are not sealed, moisture can enter the manifold which can damage the
instrument and void the warranty.
12. Make the necessary bends in the tubing to ease installation of the tubing into the output fitting
on the sample conditioning module and the stream input into the input on the feed-through
assembly.
13. Insert the tubing with the ferrule into the output fitting on the sample conditioning module.
Move the nut down onto the ferrule, screw onto the fitting and tighten.
14. Remove the plastic caps from the restrictor coils, the sealing screws from the feed-through
column vents and the sealing screw from the sample vent lines.
15. Purge the air from the transport tubing by opening the shut-off valve located on the sample
probe.
16. Insert the tubing with the ferrule into the corresponding input port located on the feed-through
assembly. Move the nut down onto the ferrule, screw into the port and tighten. Do not over-
tighten.
21 01 51 0MN A H | NG C 82 0 0 A N D P GC 10 00 US ER MA NUA L | 3 5