Figure 6-28: Save initial configuration
2.
Click Send and then Next to move to the next screen.
6.5.6
Analyze the calibration stream
This procedure determines if the analyzer provides accurate composition analysis of the calibration
gas blend. If composition analysis does not reflect the expected composition in the blend, the analyzer
requires calibration. Perform this procedure prior to running the process streams.
To run the analyzer's calibration stream:
1.
At the Analyzer Cal Stream tab, select Stream 4 on the left side of the screen
The on-screen indicator next to Stream 4 should light up, the cycle clock begins, and you
should see on-screen animation of gas flowing on the calibration stream
6 4 | N G C8 20 0 A N D PG C 10 0 0 US ER MA N UA L | 21 01 51 0MN A H
(Figure
6-29).
(Figure
6-30).