Message Reference Interpretation

IEEE-488.2 Common Commands

*CLS
Description
Examples
*ESE
Description
Parameters
Examples
*ESE?
Description
Response
Examples
*ESR?
Description
Response
Examples
*IDN?
Description
Response
Examples
Clear the Status Byte Register and all event registers, and clear the Error
Queue.
*CLS
Set the Standard Event Status Enable register.
< NR1 (0 to 255)>
*ESE 48
Query Standard Event Status Enable register.
<NR1 (0 to 255)>
*ESE?
48
Query Standard Event Status register.
<NR1 (0 to 255)>
*ESR?
48
Query the user defined instrument information.
<Manufacturer>,<Model>,<Serial number>,<Instrument firmware
version>,<DSP version>,<FPGA version>,<Internal switch
version>,<External switch version>
*IDN?
Binary-weighted decimal value. Each bit
enables or disables the appropriate bit in the
Standard Event Status register.
Set bit 4 and 5 to one and all other bits to zero.
Bit 4 and 5 are one and all other bits are zero.
Bit 4 and 5 are one and all other bits are zero.
Retrieves the IDN string which is composed by
user defined manufacturer, user defined model,
user defined serial number, instrument firmware
version, DSP version, FPGA version, internal
switch version and External switch version.