Comparator - Fluke BT5300 Series User Manual

Battery tester
Table of Contents

Advertisement

If the 300 mA measurement current is used, the Instrument will be more sensitive
to the eddy current of the measuring loop circuit. If the test can't be completed
with a large measurement current, you can reduce the effects of eddy current on
test cables, or the system refer to
Suggested
Solution, or reduce the measurement current, to complete the test.

Comparator

The Product provide a comparator function which compares measured values to user definable upper
and lower thresholds, and then display a prompt on the screen, and can also give audio prompt.
To set the comparator function:
1. In the Instrument Setup screen, use the up and down arrow (M N) on the front panel to move
the cursor to the Comparator option (the option is highlighted in yellow). Press D (Edit) to
enter the Comparator screen.
2. Use the up and down arrow (M N) keys to select the Comparator option. Press D (Edit)
button on the front panel.
3. Use the up/down arrow (M N) buttons on the front panel to select ON, then press D (OK); to
switch the comparator off, use the up/down arrow (M N) buttons on the front panel to select OFF,
then press D (OK).
4. Use the up/down arrow (M N) buttons on the front panel to move the cursor to DCV Upper
Threshold or DCV Lower Threshold (the option is highlighted in yellow), then press D
(Edit).
5. Use the up/down arrow (M N) buttons on the front panel to select DCV Upper Threshold option,
then use the numeric keypad to enter a voltage value, in V, in the DCV Upper Threshold editbox,
which can include a decimal point. As shown in the
When entering a number in the editbox, press C (Backspace) to delete the last
digit and move the cursor to the lass position.
Note
Appendix 3. Effect of Eddy Currents and
Figure 25. Comparator setting
Figrue
Note
26.
Battery Tester
Instrument Setup
51

Advertisement

Table of Contents
loading

Table of Contents