Page 1
Volt/mA Calibrator Instruction Sheet Introduction The Fluke 715 Volt/mA Calibrator is a source and measurement tool for 0 to 24 mA current loop testing and dc voltage from 0 to 10 V. The calibrator does not source and measure simultaneously.
Page 2
Contact your Fluke distributor for information about accessories. To order replacement parts or spares, see “Replacement Parts.” To contact Fluke, call one of the following telephone numbers: USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181...
Safety Information Use the calibrator only as specified in this sheet, otherwise the protection provided by the calibrator may be impaired. A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test.
Measuring dc Volts INPUT mV or V depending on range Press to toggle Press so that 100 mV and INPUT is on 10V ranges the display it01i.eps Sourcing dc Volts OUTPUT mV or V depending on range Press to toggle 100 mV and Press so that 10V ranges...
Page 5
Measuring dc mA INPUT Press to toggle mA and % of 4-20 mA scale Press so that INPUT is on the display it03i.eps...
Measuring dc mA with Loop Power INPUT Press to toggle mA and % of 4-20 mA scale Press so that INPUT is on the display +24V output Two-wire transmitter it06i.ep...
Using the Current Output Modes The calibrator provides current output in mA or percent display. Percent is -25.00 to 125.00%, where 0% is 4 mA, and 100% is 20 In source mode, the calibrator supplies the current. In simulate mode, the calibrator simulates a two-wire transmitter in an externally-powered current loop.
Page 8
Simulating a Transmitter Use simulate mode when an external 24 to 30 V loop power supply is available. Insert the test leads into the mA SIMULATE − and + jacks as shown below. OUTPUT Press to toggle mA and % of 4-20 mA scale Press to step up/down 4 mA (25%)
Review this sheet to make sure you are using the correct jacks and pushbuttons. If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement below for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee.
Page 10
Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center. Replacing the Battery Warning...
Replacing a Fuse ! Warning To avoid personal injury or damage to the calibrator, use only a 0.125A 250 V fast fuse, ® Littelfuse 2AG. Fuse F1 is probably blown if: • in the mA input mode, the calibrator always reads 0.000, even with a signal applied.
Replacement Parts and Accessories Replacement Parts Item Description PN or Qty. Model 9V battery, ANSI/NEDA 1604A or 614487 IEC 6LR61 CG81Y Holster, Yellow CG81Y F1, F2 Fuse, 125 mA, 250 V fast 686527 MP85 Case top 620200 MP86 Case bottom 620168 H2, 3, 4 Case screw...
Page 14
MP85 MP86 H2, 3, 4 MP89, 90 H5, 6 MP92 Instruction AC70A Sheet Alligator Clips TL75 Test Lead Set Holster it10c.eps...
Specifications Specifications are based on a one year calibration cycle and apply from +18°C to +28°C unless stated otherwise. “Counts” means number of increments or decrements of the least significant digit. DC V Input and Output Range Resolution Accuracy, ±(% of Reading + Counts) 100 mV 0.01 mV 0.02% + 2...
General Specifications Maximum voltage applied between any jack and earth ground or between any two jacks: 30 V Storage temperature: -40°C to 60°C Operating temperature: -10°C to 55°C Operating altitude: 3000 meters maximum Temperature coefficient: ±0.005% of range per °C for the temperature ranges -10 to 18°C and 28 to 55°C Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up to 55°C...
Page 17
LIMITED WARRANTY & LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling.