Fluke 726 User Manual

Fluke 726 User Manual

Multifunction process calibrator
Hide thumbs Also See for 726:
Table of Contents

Advertisement

®
726
Multifunction Process Calibrator
Users Manual
September 2005
© 2005 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 726

  • Page 1 ® Multifunction Process Calibrator Users Manual September 2005 © 2005 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
  • Page 2 LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction........................1 Contacting Fluke......................1 Standard Equipment...................... 3 Safety Information ......................3 Symbols ........................7 Getting Acquainted with the Calibrator ................8 Input and Output Terminals ..................8 Keys.......................... 10 Display ........................13 Configuration Menus ..................... 14 Contrast Adjustment ....................
  • Page 4 Users Manual  HART Resistor ON/OFF ..................16 Getting Started ....................... 16 Voltage to Voltage Test ..................... 16 Using Measure Mode ..................... 18 Measuring Electrical Parameters (Upper Display) ............ 18 Current Measurement with Loop Power ..............18 Measuring Electrical Parameters (Lower Display) ............ 20 Measuring Temperature ....................
  • Page 5 Cleaning the Calibrator ..................... 54 Service Center Calibration or Repair ................ 54 Replacement Parts ....................54 Accessories ........................56 External Fluke Pressure Module Compatibility ............56 Specifications ........................ 59 DC Voltage Measurement and Source ..............59 DC mA Measurement and Source ................59 Ohms Measurement ....................
  • Page 6 Users Manual Loop Power Supply ....................64 Pulse Read and Pulse Source .................. 64 Pressure Measurement ..................... 64 General Specifications ....................65 Index...
  • Page 7 International Symbols ......................7 Input/Output Terminals and Connectors................9 Key Functions ........................11 Thermocouple Types Accepted .................... 22 RTD Types Accepted......................25 mA Step Values ........................42 Replacement Parts ....................... 54 Fluke Pressure Module Compatibility ................... 56 Pressure Modules......................... 57...
  • Page 8 Users Manual...
  • Page 9: List Of Figures

    List of Figures Figure Title Page Standard Equipment ......................6 Input/Output Terminals and Connectors................8 Keys............................10 Elements of a Typical Display ....................13 Adjusting the Contrast ......................14 Voltage-to-Voltage Test ......................17 Measuring Voltage and Current Output ................18 Connections for Supplying Loop Power................
  • Page 10 Users Manual Connections for Sourcing Pressure ..................40 SAVE DATA Menu Showing Measurement Memory Location 3, 1 ........44 Calibrating a Thermocouple Transmitter ................46 Calibrating a Pressure-to-Current (P/I) Transmitter .............. 48 Calibrating a Current-to-Pressure (I/P) Transmitter .............. 50 Calibrating a Chart Recorder ....................52 Replacing the Batteries ......................
  • Page 11: Introduction

    Multifunction Process Calibrator Stores and recalls calibration screens. Introduction Control the Calibrator remotely from a PC running a The Fluke 726 Multifunction Process Calibrator (referred terminal emulator program. to as “the Calibrator”) is a handheld, battery-operated Contacting Fluke instrument that measures and sources electrical and physical parameters.
  • Page 12: Summary Of Source And Measure Functions

    Users Manual Table 1. Summary of Source and Measure Functions Function Measure Source dc V 0 V to 30 V 0 V to 20 V dc mA 0 to 24 mA 0 to 24 mA Frequency 2 CPM to 15 kHz 2 CPM to 15 kHz Resistance to 4000...
  • Page 13: Standard Equipment

    Stackable alligator clip test leads provided by the Calibrator may be impaired. 726 Product Overview (not shown in Figure 1) A Warning identifies conditions and actions that pose hazard(s) to the user. A Caution identifies conditions and 725/726 CD-ROM (contains Users Manual;...
  • Page 14 Users Manual XW Warning To avoid possible electric shock or personal injury: Use the Calibrator only as described in the Users Manual or the protection provided by the Calibrator may be impaired. Do not apply more than the rated voltage, as marked on the Calibrator, between the terminals, or between any terminal and earth ground (30 V 24 mA max all terminals).
  • Page 15 Multifunction Process Calibrator Safety Information When using a pressure module, make sure the process pressure line is shut off and depressurized before connecting it or disconnecting it from the pressure module. Use only 4 AA batteries, properly installed in the Calibrator case, to power the Calibrator. Disconnect test leads before changing to another measure or source function.
  • Page 16 Users Manual AC72 Alligator Clips TL75 Test Lead Set Stackable Test Leads bec01f.eps Figure 1. Standard Equipment...
  • Page 17: Symbols

    Multifunction Process Calibrator Safety Information Symbols Symbols used on the Calibrator and in this manual are explained in Table 2. Table 2. International Symbols AC - Alternating current Double insulated DC - Direct current Battery Risk of danger. Important information. See Earth ground Manual.
  • Page 18: Getting Acquainted With The Calibrator

    Users Manual Getting Acquainted with the Calibrator Input and Output Terminals Figure 2 shows the Calibrator input and output terminals. Table 3 explains their use. PRECISION CALIBRATOR ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH LOOP TEST HART V mA FREQ MEAS SOURCE...
  • Page 19: Input/Output Terminals And Connectors

    Multifunction Process Calibrator Getting Acquainted with the Calibrator Table 3. Input/Output Terminals and Connectors Name Description Pressure module Connects the Calibrator to a pressure module or to a PC for a remote control connector/serial connector serial connection. MEASURE V, mA terminals Input terminals for measuring voltage, current, supplying loop power, HART B, C resistance, switch test options.
  • Page 20: Keys

    Users Manual Keys Figure 3 shows the Calibrator keys and Table 4 explains their use. PRECISION CALIBRATOR PRECISION CALIBRATOR ZERO ZERO %Error OPEN/CLOSE %Error OPEN/CLOSE 3 Seconds 3 Seconds V mA V mA SWITCH SWITCH LOOP TEST LOOP TEST HART HART V mA FREQ...
  • Page 21: Key Functions

    Multifunction Process Calibrator Getting Acquainted with the Calibrator Table 4. Key Functions Name Description Turns the power on or off. %Error Toggles voltage, mA, or Loop Power and % Error measurement functions in the upper display. ZERO Selects the pressure measurement function in the upper display. Repeated pushes cycle through the 3 Seconds different pressure units.
  • Page 22 Users Manual Table 4. Key Functions (cont.) Name Description Cycles through : E Slow repeating 0 % - 100 % - 0 % ramp TRIGGER/STOP P Fast repeating 0 % - 100 % - 0 % ramp N Repeating 0 % - 100 % - 0 % ramp in 25 % steps Used for the pulse train and totalizer functions.
  • Page 23: Display

    Multifunction Process Calibrator Getting Acquainted with the Calibrator Display Figure 4 shows the elements of a typical display. Memory Locations Loop HART Mode for Calibrator Setups Indicator Indicator Low Battery Symbol Units Display Mode Indicator % Error Indicators Auto Ramp bec07f.eps Figure 4.
  • Page 24: Configuration Menus

    Users Manual Configuration Menus Use the configuration menus to set or change these parameters of the Calibrator: PRECISION CALIBRATOR Contrast Adjustment Shut Down Mode CJC on/off C/ F ZERO OPEN/CLOSE %Error 3 Seconds V mA SWITCH LOOP TEST Frequency/Pulse output voltage HART Pulse output frequency V mA...
  • Page 25: Shut Down Mode

    Multifunction Process Calibrator Configuration Menus Shut Down Mode Frequency Pulse Output Voltage Press C until FREQ OUTPUT V Adjust appears The Calibrator comes with a shut down mode set for 30 minutes (displayed for about 1 second when the on the display. Calibrator is initially turned on).
  • Page 26: Hart  Resistor On/Off

    Users Manual  Resistor ON/OFF Getting Started HART Press C until SELECT HART ON or OFF This section details some basic operations of the appears on the display. Calibrator. Use V to toggle ON or OFF. Voltage to Voltage Test Press S to save the setting.
  • Page 27 Multifunction Process Calibrator Getting Started bec39f.eps Figure 6. Voltage-to-Voltage Test...
  • Page 28: Using Measure Mode

    Users Manual Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, or to measure the output of a 700 Series pressure module, use the upper display and proceed as follows: Press to select volts or current.
  • Page 29 Multifunction Process Calibrator Using Measure Mode PRECISION CALIBRATOR ZERO OPEN/CLOSE %Error 3 Seconds V mA SWITCH LOOP TEST HART V mA FREQ MEAS SOURCE PULSE TEST DC PWR – + – EXIT CONFIG CONFIG 100% SELECTION SAVE RECALL ENTER – TRIGGER/STOP Return to Recall...
  • Page 30: Measuring Electrical Parameters (Lower Display)

    Users Manual Measuring Electrical Parameters (Lower Display) To measure electrical parameters using the lower display, PRECISION CALIBRATOR proceed as follows: Connect the Calibrator as shown in Figure 9. If necessary, press for MEASURE mode (lower display). Press for dc voltage or current, ZERO OPEN/CLOSE %Error...
  • Page 31: Measuring Temperature

    Multifunction Process Calibrator Using Measure Mode W Caution Measuring Temperature One pin is wider than the other. Do not try to force a Using Thermocouples miniplug into the wrong polarization. The Calibrator supports 13 standard thermocouples. Note Table 5 summarizes the ranges and characteristics of each.
  • Page 32: Thermocouple Types Accepted

    Users Manual Table 5. Thermocouple Types Accepted Positive Lead (H) Type Positive Lead Color Negative Lead Specified Range Material Material ( C) ANSI* IEC** Chromel Purple Violet Constantan -200 to 950 Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300 Iron White Black Constantan -200 to 1200...
  • Page 33 Multifunction Process Calibrator Using Measure Mode 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + Process Temperature LOOP mA - Warning 30 V maximum to TC Miniplug bec12f.eps Figure 10. Measuring Temperature with a Thermocouple...
  • Page 34: Using Resistance-Temperature Detectors (Rtds)

    PRT Custom Curves Up to three custom curves can be named and CVD coefficients can be entered through the serial port. Names can be up to six characters. For more information, see the Application Note on the 725/726 CD.
  • Page 35: Rtd Types Accepted

    Multifunction Process Calibrator Using Measure Mode Table 6. RTD Types Accepted RTD Type Ice Point (R Material Range ( C) Pt100 (3926) Platinum 0.003926 -200 to 630 Pt100 (385) Platinum 0.00385 -200 to 800 Ni120 (672) Nickel 0.00672 -80 to 260 Pt200 (385) Platinum 0.00385...
  • Page 36 Users Manual 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA - 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA - 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA - bec15f.eps Figure 11.
  • Page 37: Measuring Pressure

    Using Measure Mode Measuring Pressure Many ranges and types of pressure modules are available Gage Differential from Fluke, see “Accessories”. Before using a pressure High module, read its instruction sheet. The modules vary in 700 P 06 700 P 04...
  • Page 38: Zeroing With Absolute Pressure Modules

    Users Manual Never apply pressure above the rated Zeroing with Absolute Pressure Modules maximum printed on the pressure To zero, adjust the Calibrator to read a known pressure. module. This can be barometric pressure, if it is accurately known, Only use the pressure module with for all but the 700PA3 module.
  • Page 39 Multifunction Process Calibrator Using Measure Mode Isolation Valve Gage Module PRECISION CALIBRATOR ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH TEST LOOP HART V mA FREQ MEAS SOURCE PULSE EXIT CONFIG Differential CONFIG 100% SELECTION Module SAVE RECALL ENTER TRIGGER/STOP Tank Return to Recall...
  • Page 40: Using Source Mode

    Users Manual Using Source Mode Simulating a 4- to 20-mA Transmitter Simulate is a special mode of operation in which the In SOURCE mode, the Calibrator: Calibrator is connected into a loop in place of a generates calibrated signals for testing and transmitter and supplies a known, settable test current.
  • Page 41 Multifunction Process Calibrator Using Source Mode PRECISION CALIBRATOR ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH LOOP TEST HART V mA FREQ MEAS SOURCE PULSE EXIT CONFIG CONFIG 100% SELECTION +24 V SAVE RECALL ENTER TRIGGER/STOP Return to Recall Loop Power Black Supply...
  • Page 42: Sourcing Other Electrical Parameters

    Users Manual Press for dc voltage, or for frequency, and Sourcing Other Electrical Parameters R for resistance. Volts, ohms, and frequency are also sourced and shown Enter the desired output value by pressing in the lower display. keys. Press to select a different digit to To select an electrical sourcing function, proceed as change.
  • Page 43 Multifunction Process Calibrator Using Source Mode 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE + mA mA + LOOP mA - HART – Communicator Black Common 30V MAX ALL TERMINALS SOURCE / MEASURE MEASURE mA + LOOP mA - – Black Common bec16f.eps...
  • Page 44: Simulating Thermocouples

    Users Manual Simulating Thermocouples Figure 16 shows this connection. Proceed as follows to simulate a thermocouple: Connect the Calibrator TC input/output to the instrument Attach the thermocouple leads to the appropriate TC under test with the thermocouple wire and the appropriate miniplug, then to the TC input/output as shown in thermocouple mini-connector (polarized thermocouple Figure 16.
  • Page 45 Multifunction Process Calibrator Using Source Mode PRECISION CALIBRATOR TEST DC PWR – Color depends on type of TC ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH TEST LOOP – HART V mA FREQ MEAS SOURCE PULSE EXIT CONFIG CONFIG 100% SELECTION SAVE RECALL...
  • Page 46: Simulating Rtds

    Connect the Calibrator to the instrument under test as If the 726 display indicates ExI HI, the excitation shown in Figure 17. Proceed as follows to simulate an current from your device under test exceeds the RTD: limits of the 726.
  • Page 47 Multifunction Process Calibrator Using Source Mode SENSOR TERMINALS BLACK bec40f.eps Figure 17. Connections for Simulating 3- and 4-Wire RTD...
  • Page 48: Sourcing Pressure

    SOURCE field. Figure 18 shows how to Never apply more than 10 ft.-lb. (13.5 Nm) connect a pump to a Fluke pressure module which makes of torque between the pressure module it a calibrated source.
  • Page 49 Multifunction Process Calibrator Using Source Mode Connect a pressure module to the Calibrator as Zero the pressure module as described in the shown in Figure 18. The threads on the pressure module’s Instruction Sheet. Modules vary in zeroing modules accept standard ¼ NPT pipe fittings. Use procedures depending on module type.
  • Page 50 Users Manual Hand Pump Measure mA Source Pressure 24 V Loop Power PRECISION CALIBRATOR Enabled ZERO %Error OPEN/CLOSE Pressure 3 Seconds V mA SWITCH TEST LOOP Module HART V mA FREQ To Unit Under Test MEAS SOURCE PULSE EXIT CONFIG CONFIG 100% SELECTION...
  • Page 51: Setting 0 % And 100 % Output Parameters

    Multifunction Process Calibrator Setting 0 % and 100 % Output Parameters % Error Functionality Setting 0 % and 100 % Output Parameters Percentage Error is available for every range on the lower display. The calculation is based on a mA For current output, the Calibrator assumes that 0 % percentage deviation from the value measured on the corresponds to 4 mA and 100 % corresponds to 20 mA.
  • Page 52: Manually Stepping The Ma Output

    Users Manual Manually Stepping the mA Output To exit ramping, press any button. Table 7. mA Step Values To manually step current output: to step the current up or down in Step 4 to 20 mA 25 % steps. 4.000 Touch either to go to 0 %, or to go to...
  • Page 53: Recall A Setup

    Multifunction Process Calibrator Storing and Recalling Data Recall a Setup Storing and Recalling Data To recall a setup: Up to 40 data samples can be stored in a nonvolatile memory to be recalled for later use. A low battery Push S twice. The right side of the display condition or a battery change does not jeopardize the changes to show RECL SETUP and RECALL DATA.
  • Page 54: Recall Data

    Users Manual Push Y or Z to select the correct data location Memory Locations Save Data Mode Indicator (bottom right of display). Push S to recall the data stored in that location. Data Point S A V E Push W to return to the same RECALL DATA D A T A location to see the next saved measurement, 2 of 5 for example.
  • Page 55: Calibrating A Transmitter

    Multifunction Process Calibrator Calibrating a Transmitter Calibrating a Transmitter If necessary, press for SOURCE mode. Set the zero and span parameters by pressing Use the measurement (upper display) and source (lower . Enter these parameters by pressing and display) modes to calibrate a transmitter. This section holding .
  • Page 56 Users Manual PRECISION CALIBRATOR ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH LOOP TEST HART V mA FREQ MEAS SOURCE PULSE TEST DC PWR – + – EXIT CONFIG CONFIG 100% SELECTION SAVE RECALL ENTER – TRIGGER/STOP Return to Recall HART Communicator Black...
  • Page 57: Calibrating A Pressure Transmitter

    Multifunction Process Calibrator Calibrating a Pressure Transmitter Calibrating a Pressure Transmitter If necessary, press for SOURCE mode. Zero the pressure module. The following steps explain calibrating a pressure transmitter. Perform checks at 0 % and 100 % of span and adjust the transmitter as necessary.
  • Page 58 Users Manual Hand Pump S I G N A L Measure mA – Source Pressure 24 V Loop Power PRECISION CALIBRATOR Enabled T E S T ZERO %Error OPEN/CLOSE Pressure 3 Seconds V mA SWITCH TEST LOOP Module HART V mA FREQ MEAS SOURCE...
  • Page 59: Calibrating An I/P Device

    Multifunction Process Calibrator Calibrating an I/P Device Calibrating an I/P Device Press for sourcing current (lower display). If necessary, press for SOURCE mode. The following steps explain how to calibrate a device that controls pressure. Proceed as follows: Enter the desired current by pressing Press to select different digits.
  • Page 60 Users Manual S I G N A L Measure – Pressure Source mA PRECISION CALIBRATOR T E S T Pressure ZERO %Error OPEN/CLOSE 3 Seconds V mA SWITCH Module TEST LOOP HART V mA FREQ MEAS SOURCE PULSE EXIT CONFIG CONFIG 100% SELECTION...
  • Page 61: Pressure Switch Test

    Multifunction Process Calibrator Pressure Switch Test Pressure Switch Test OPEN displays once the switch is open. Slowly bleed the pump until the pressure switch closes. RECALL appears Note on the display. This example uses a normally-closed switch. Press Q to read the pressure values for when the The procedure is the same for an open switch switch opened, for when it closed, and for the but the display reads OPEN instead of CLOSE.
  • Page 62: Remote Control Commands

    The Calibrator may be remotely controlled from a PC running a terminal emulator program. The remote control commands give access to all capabilities of the Calibrator with the exception of pressure measurement. See the Fluke Web Site for the 726 Remote Programming application note at www.fluke.com/processtools ...
  • Page 63: Maintenance

    Multifunction Process Calibrator Maintenance Maintenance Replacing the Batteries Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator (M) appears. Figure 24 shows you how to replace the batteries. Battery and Compartment bec38f.eps...
  • Page 64: Cleaning The Calibrator

    659026 and replace them if needed. Battery door 1/4-turn 948609 To locate an authorized service center, refer to fasteners “Contacting Fluke” at the beginning of the manual. TL75 series test leads 855742 Test lead, red 688051 Test lead, black 688066...
  • Page 65 Multifunction Process Calibrator Maintenance bec45f.eps Figure 25. Replacement Parts...
  • Page 66: Accessories

    The output of Fluke 700P pressure modules can cause your Fluke representative. Fluke Pressure Module the 726’s 5 digit display to overflow, or else produce Compatibility is listed in Table 9. Pressure Modules and values that are too low to be read if inappropriate units Fluke model numbers are listed in Table 10.
  • Page 67: Pressure Modules

    Multifunction Process Calibrator Accessories Table 10. Pressure Modules Fluke Model Number Range Type and Media Fluke-700P00 0 to 1” H differential, dry Fluke-700P01 0 to 10” H differential, dry Fluke-700P02 0 to 1 psi differential, dry Fluke-700P22 0 to 1 psi...
  • Page 68 Users Manual Table 10. Pressure Modules (cont.) Fluke Model Number Range Type and Media Fluke-700P29 0 to 3,000 psi gage, wet Fluke-700P30 0 to 5,000 psi gage, wet Fluke-700P31 0 to 10,000 psi gage, wet Fluke-700PA3 0 to 5 psi...
  • Page 69: Specifications

    Multifunction Process Calibrator Specifications Specifications Specifications are based on a one year calibration cycle and apply from +18 C to +28 C unless stated otherwise. All specifications assume a five-minute warmup period. DC Voltage Measurement and Source Accuracy, Range Minimum Maximum (% of Reading + Floor) 30 V (upper display)
  • Page 70: Ohms Measurement

    Users Manual Ohms Measurement Accuracy Ohms Range Minimum Maximum (% of Reading + Floor) Ohms Read (low) 0.00 400.00 0.015 % + 0.05 Ohms Read (high) 401.0 4000.0 0.015 % + 0.5 Ohms Source Excitation Current from Accuracy Ohms Range Minimum Maximum Measurement Device...
  • Page 71: Frequency Source

    Multifunction Process Calibrator Specifications Frequency Source Range Minimum Maximum Accuracy CPM Source 1000 0.05 % Hz Source 1000.0 0.05 % KHz Source 10.00 0.25 % 10.00 15.00 0.50 % Temperature, Thermocouples CJC ON CJC OFF Type Minimum Maximum Accuracy Accuracy -210 1200 -200...
  • Page 72 Users Manual CJC ON CJC OFF Type Minimum Maximum Accuracy Accuracy 1767 1000 1000 1820 1000 1000 2316 -200 0.45 0.25 -200 0.45 0.25 -200 1300 -200 2500 Range Accuracy Thermocouple in mV read 0.015 % + 10 µV -10 C to 75 C (% of Reading + Floor) Thermocouple in mV source 0.015 % + 10 µV...
  • Page 73: Rtd Accuracy (Read And Source) (Its-90)

    Multifunction Process Calibrator Specifications RTD Accuracy (Read and Source) (ITS-90) Range Minimum Maximum Accuracy Ni120 (672) -80.00 260.00 0.15 -200.00 100.00 0.15 100.00 300.00 0.25 Pt100 (385) 300.00 600.00 0.35 600.00 800.00 0.45 -200.00 100.00 0.15 Pt100 (3926) 100.00 300.00 0.25 300.00 630.00...
  • Page 74: Loop Power Supply

    Users Manual Loop Power Supply Voltage: 24 V Maximum current: 22 mA Short circuit protected. Pulse Read and Pulse Source Pulse Accuracy Frequency 10,000 1 Count 2 CPM to 10 kHz Source 100,000 Read Pressure Measurement Range Resolution Accuracy Units Mode Determined by 5 digits...
  • Page 75: General Specifications

    Multifunction Process Calibrator Specifications General Specifications Operating temperature -10 C to 50 C Storage temperature - 20 C to 70 C Stability 0.005 % of range/ C outside of 23 Operating altitude 3000 meters above mean sea level Relative Humidity (% RH operating without condensation) 90 % (10 to 30 C) 75 % (30 to 40 C) 45 % (40 to 50 C)
  • Page 76 Users Manual...
  • Page 77: Index

    Index —0— —B— —D— 0% output parameter, setting, 41 Battery, replacing, 53 Data Recall, 44 display —1— —C— contrast adjustment, 14 Display, 13 100% output parameter, setting, 41 Calibration, 54 Celcius and Fahrenheit, 15 Cleaning calibrator, 54 —E— —4— Cold Junction Compensation (CJC), 15 Commands Electrical parameters 4 to 20 mA transmitter...
  • Page 78 Users Manual —G— —M— Pressure modules, zeroing, 28 Pressure Switch Test, 51 Getting started, 16 Measure functions, summary (table), 2 Pressure transmitter, calibrating, 47 Measure mode, 18 Pulse Train Source/Read, 44 Measuring —H— pressure, 27 —R— temperature with RTDs, 24 HART temperature with thermocouples, 21 configuration menu, 16...
  • Page 79 (continued) Index storing, 42, 43 input, 8 shut down mode, 15 output, 8 Thermocouple, 34 Thermocouple Simulating measuring, 21 loop power, 30 measuring temperature, 21 RTD, 36 sourcing, 34 thermocouples, 34 types, 21 Source functions, summary (table), 2 thermocouples Source mode, 30 accepted types, 22 Sourcing Transmitter...
  • Page 80 Users Manual...

Table of Contents