Figure 59. Schematic diagram of switching circuit - front panel
In
Figure
59, SOURCE HI, SOURCE LO, SENSE HI and SENSE LO correspond to the front panel
input terminals of BT5321; K1, K2, K3 and K4 are relays, numbers 1~6 represent probe tips, and line
segments represent connected test cables.
1. For different test functions, it is necessary to set the relays to the corresponding state first:
Test function
ACR+DCV
Enclosure potential
contact check
Positive to
enclosure voltage
Negative to
enclosure voltage
2. After the warming up of the BT5321, complete different measurement according to the following
instructions.
1) Initializing instrument settings
*RST
SYSTem:CALibration
2) ACR+DCV measurement function: measure the internal resistance and open-circuit voltage
of the cell
FUNCtion RVOLtage
RESistance:RANGe 0.003
146
K1 state
Down
Up
Down
Up
Command
Command
K2 state
K3 state
Down
Down
Up
Up
Up or down
Down
Reset the Instrument
Execute one self-calibration. See the section
Calibraion
for more details.
Set the function to ACR+DCV measurement.
Select a appropriate resistance range, such as 3 mΩ. The
voltage range does not to be set since there is only one
K4 state
Close
Up
Close
Open
Action
Self-
Action