Command
FUNCtion RVOLtage
INITiate:CONTinuous OFF
INITiate
STATus:OPERation?
FETCh?
Module Internal, One-Shot Scan, External Trigger
In this scenario it is desired to scan and measure several internal switch channels in a one-shot
fashion under external triggering.
Command
*RST
*CLS
TRIGger:SOURce EXTernal
SWITch:MODule INTernal
SWIT:MOD:STAT? INT
ROUTe:SCAN (@101:132)
FUNCtion RVOLtage
INITiate:CONTinuous OFF
INITiate
STATus:OPERation?
*TRG
STATus:OPERation?
FETCh?
Configure to measure the ACR+DCV.
Disable continuous measurement.
Initiate scan.
Query the Operation Statues register until the Bit4 (Sweep
Done) and Bit 8 (Scan Done) is set.
Return the latest reading (32 pairs of voltage and resistance).
Reset the instrument to the power-on state.
Clear the Status Byte Register and all event registers and clear
the Error Queue.
Configure the trigger source to be external.
Configure the instrument to use the internal switch channel as
measurement input channel.
Query the internal switch module's slots state.
According to the active slots' state, configure scan channel list
from 101 to 132.
Configure to measure the ACR+DCV.
Disable continuous measurement.
Initiate scan.
Query the Operation Status register, until the Bit 12 (Ready for
Initiate Trigger) is set.
Trigger one measurement (Trigger key signal can achieve the
same result).
Repeat the last 2 steps until the Bit 4 (Sweep Done) and Bit 8
(Scan Done) of the Operation Status register is set.
Return the latest reading (32 pairs of voltage and resistance).
Battery Tester
Remote Control
Action
Action
115