Integral T process thermostats
after approx. 2 s
after approx. 2 s
21/01/2019 / YAWE 0026
Calibrating the inputs
Apply the voltage or current for the appropriate range
limits (0 V, 10 V, 0 mA, 20 mA). Deviations up to
approx. 10 % of range can be corrected.
The calibration values remain stored!
Factory calibration is performed at 0 V, 10 V, 0 mA and
20 mA.
Feed in current 0 mA at pins 4 (+) and 3 (-) on
connector 66S.
Feed in current 20 mA.
Calibration of setpoint input current is completed.
Setpoint input voltage and external temperature input
current and voltage are calibrated similarly.
Calibrating the outputs
Connect up the device to be supplied or an accurate
multimetre with a current range 0 - 20 mA or a voltage
range 0 - 10 V.
Calibrate the outputs with the required termination
resistance.
Starting up
71