Integral T process thermostats
21/01/2019 / YAWE 0026
MAX = maximum temperature in °C.
For determining the temperature range to which the
current or voltage range has to be assigned.
Example: 0 – 10 V is to correspond to –50 – 150 °C
MIN = -50; MAX = 150. At the setpoint input only
values in the range –30 °C to 120 °C are processed
since this corresponds to the operating range of the
equipment. Outside this range the display shows the
limit value.
Back to AIS.
Scroll to submenu "Analogue inputs", but referred to
the configuration and scaling of the external
temperature input ( Section 7.6.5.1).
Continue as for setpoint input AIS.
Error message, in the current range 4 - 20 mA the
current is less than 4 mA (0 to 4 mA).