Lauda Integral T 1200 Operating Instructions Manual page 74

Process thermostats
Table of Contents

Advertisement

With external control switched on:
74
Integral T process thermostats
Pressing
scrolls from REG . . to the input for the
control parameters.
With internal control (CON I)
With internal control (CON I) select here the
proportional band Xp for the outflow controller. Input
range 0.1 – 10 °C. Useful values are 2 – 7 °C,
depending on external circuit and heat transfer liquid. If
the value is too small (e.g. 2 °C) this may result in
control oscillations.
If the value is too large (e.g. 8 °C) the compensation of
disturbances may be worse and more sluggish.
Starting up
YAWE 0026 / 21/01/2019