Connection of the external Pt100 to Lemo socket 10S:
Pin
1
+
2
+
3
4
4-pin Lemosa plug for Pt100 connection (Cat. No. EQS 022).
Use screened connecting cable. Connect screen to connector case.
7.5.3
Pressure indication
7.5.4
Menu . .
36
I
Current path
U
Voltage path
-
U
Voltage path
-
I
Current path
Integral T process thermostats
When external Pt100 is not connected.
Line 2 indicates the pump pressure which establishes
itself through the flow resistance in the external circuit.
This indication is also required for adjusting and
monitoring the bypass.
The actual pump flow rate can be deduced from the
pump characteristic.
Scroll to level 1.
Starting up
Pt100
DIN EN 60751
YAWE 0026 / 21/01/2019