Lauda Integral T 1200 Operating Instructions Manual page 22

Process thermostats
Table of Contents

Advertisement

Water-cooled version
The heat of condensation and the motor heat are dissipated through a water-cooled counter flow heat
exchanger. Connect up the hoses. The connection for inlet (from water tap) and outlet (to drain) are
located on the back of the equipment at the bottom. Inlet on the left, outlet to drain on the right, looking
on the back of the equipment. The cooling water flow is automatically adjusted to the required amount
through the condenser pressure.
Water inlet
Secure the hoses with hose clips to prevent slipping off!!
The water consumption depends on the energy to be dissipated. It ranges between 200 and 2000 L/h
depending on equipment type, cooling water temperature and loading. The cooling water temperature
must not exceed 25 °C. Cooling water pressure should be
not exceed be 10 bar.
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
Differential pressure  p (feed - outlet)
Cooling water temperature
Cooling water quantity
Cooling water hose for connection to the
device
22
Integral T process thermostats
2.5 bar. The maximum overpressure must
>
maximum 10 bar overpressure
minimum 2.5 bar
10 to 15 °C recommended,
10 to 25 °C not exceed (with power restrictions)
see Technical Data ( 9)
minimum 13 mm
Preparations
Water outlet
YAWE 0026 / 21/01/2019