Sensor Temperature Unit

Fast Keys
1, 4, 1, 2, 2
DIAGNOSTICS AND
SERVICE

Master Reset

Fast Keys
Field Communicator
To configure the process alerts with a Field Communicator, perform the
following procedure:
1.
From the HOME screen, follow the fast key sequence "Alerts." (1
Device Setup, 3 Configuration, 4 Device Output Configuration, 4
Alerts)
2.
Select 1 for HI-HI Alert.
Select 2 for HI Alert.
Select 3 for LO Alert.
Select 4 for LO-LO Alert.
3.
Select 1 to configure the alert.
AMS
Right click on the device and select "Configure/Setup" from the menu
1. On the "Alerts" Page, press the "Configure Alert" button for each of the
process alerts desired (1 per tab). Follow the prompts from the alert
configuration wizard.
The Sensor Temperature Unit command selects between Celsius and
Fahrenheit units for the sensor temperature. The sensor temperature output
is accessible via HART only.
Field Communicator
Enter the fast key sequence "Sensor Temperature Unit."
AMS
Right click on the device and select "Configure/Setup" from the menu
1. In the "Temperature" tab, use the drop down menu "Snsr temp unit" to
select F (Farenheit) or C (Celsius). Click Apply.
2. Click Next to acknowledge send warning.
3. Select Finish to acknowledge the method is complete.
4. After carefully reading the warning, select yes.
Diagnostics and service functions listed below are primarily for use after field
installation. The Transmitter Test feature is designed to verify that the
transmitter is operating properly, and can be performed either on the bench or
in the field.
The master reset will reset device electronics.
1, 2, 1
Field Communicator
Enter the fast key sequence "Master Reset."
AMS
Right click on the device and select "Methods>Diagnostics and Test," then
"Master Reset" from the menu
Rosemount 3051S Wireless Series
.
.
.
2-9