Rosemount 305 Integral
Manifold Installation
Procedure
Rosemount 3051S Wireless Series
The Rosemount 305 is available in two designs: Traditional and Coplanar.
The traditional 305 Integral Manifold can be mounted to most primary
elements with mounting adapters in the market today. The Rosemount 306
Integral Manifold is used with In-line transmitters to provide block-and-bleed
valve capabilities of up to 10000 psi (690 bar). The Rosemount 304 comes in
two basic styles: traditional (flange x flange and flange x pipe) and wafer. The
304 traditional manifold comes in 2, 3, and 5-valve configurations. The 304
wafer manifold comes in 3 and 5 valve configurations.
CONVENTIONAL
To install a 305 Integral Manifold to a 3051S Wireless transmitter:
1. Inspect the Teflon SuperModule O-rings. If the O-rings are undamaged,
reusing them is recommended. If the O-rings are damaged (if they have
nicks or cuts, for example), replace them with new O-rings.
IMPORTANT
If replacing the O-rings, take care not to scratch or deface the O-ring grooves
or the surface of the isolating diaphragm while you remove the damaged
O-rings.
LCD Display
Meter
Cover
COPLANAR
TRADITIONAL
IN-LINE
3-13