insert is facing outward with the beveled portion toward the interior of the sensor.
Then standardize the instrument with the UV filter in place.
• Sample Cup Set for Translucent Liquids
This set provides a device for holding powders, pellets, granules, and translucent
liquids at the sample port. It includes a 64 mm (2.5 inch) glass sample cup, plastic
ring and ceramic disk set, special port plate insert, and sample cup cover
The ring and disk set are provided for translucent liquids, transparent liquids, and
semi-solids where the path length is related to color concentration and must be
fixed. First, the sample cup port plate should be installed to hold the sample cup in
the proper position. Place the ring inside the sample cup and fill the cup with sample
to a level above the ring. Then place the ceramic disk on top of the sample (until it
rests on top of the ring) with the white portion facing the sample. Cover with the
sample cup cover to exclude external light. Measure the sample at the sample port
through the glass bottom of the sample cup with the instrument in the port-up
orientation. .
• Sample Cup Port Plate
This port insert is specially designed to accommodate the 64 mm (2.5 in) sample cup
and to hold it in the proper position at the measurement port. Remove the standard
port insert by pulling it out of the port insert retainer. Replace it with the sample cup
port insert, then standardize the instrument.
• 64 mm (2.5 in) Glass Sample Cup
The sample cup is ideal for the analysis of powders, granules, pellets, and
translucent samples. It is recommended for use with the instrument in the port-up
orientation. The cup can be filled with the desired sample and placed at the
measurement port so that the sample is measured through the glass bottom of the
cup.
ColorFlex EZ User's Manual Version 2.2
Figure 86. Sample Cup Port Plate
Figure 87. Glass Sample Cup
63