Stdz Interval; Autosearch; Color Diff - HunterLab ColorFlex EZ User Manual

Hide thumbs Also See for ColorFlex EZ:
Table of Contents

Advertisement

Stdz Interval

Indicate the amount of time that must elapse (in hours) after the last standardization and
before the instrument prompts for re-standardization. Press the RIGHT arrow button to
increase the number of hours or the LEFT arrow button to decrease the number. Values
between 1 and 16 may be selected in increments of 1. Selecting 4 is suggested. The choice of
OFF disables prompting for standardization, but its use is not recommended.
AutoSearch
One of the features of the ColorFlex EZ is that it can find the closest physical, numeric, or hitch
standard stored in the instrument's memory to a sample read.
When AUTOSEARCH is highlighted and the RIGHT arrow button is pressed, a submenu appears.
In this submenu, the following parameters may be set:

Color Diff

To deselect the AUTOSEARCH feature, set this parameter to (...). To use this feature,
select the color difference index for comparison of the sample/standard. The choices
are dE, dE*, and dEcmc. This index and the illuminant/observer set in each product
setup searched will be used for the comparison.
34
ColorFlex EZ User's Manual Version 2.2
Figure 49. AutoSave
Figure 50. Standardization Interval
Figure 51. Color Difference