Peep System (Test 6) - Respironics ESPRIT Service Manual

Ventilator
Hide thumbs Also See for ESPRIT:
Table of Contents

Advertisement

REF 580-1000-02 G
7. Touch Exhalation and adjust the steps until the analyzer pressure
reads 100 ± 5 cmH
8. Check that the Inhalation Pressure and Exhalation Pressure displays
on the hardware screen read within ± 10% of the analyzer display.
EXAMPLE: If the analyzer's display reads 98.7 cmH
display is 9.87 cmH
88.83 to 108.57 cmH
+ 10% = 98.7 + 9.87 = 108.57 cmH
9. The pressure accuracy test is complete.
Figure 8-20: Pressure Accuracy Test Configuration

PEEP System (Test 6)

The PEEP system test verifies the integrity of the PEEP system and checks for
autocycling.
Esprit® Ventilator Service Manual © Respironics, Inc.
Performance Verification
O (95 to 105 cmH
O).
2
2
O, and ± 10% of the analyzer display would be
2
O (-10% = 98.7 - 9.87 = 88.83 cmH
2
O).
2
Chapter 8
O, 10% of the
2
O, and
2
8-21