Chapter 8
Performance Verification
Preliminary
Pneumatic
Calibration Analyzer
Setup
Certifier FA Plus
Setup
8-6
g. Turn the ventilator OFF.
Figure 8-2: Diagnostic Mode - Software Screen
When using a pneumatic calibration analyzer, remember:
•
The analyzer should be in BTPS when measuring volume.
•
The analyzer should be in ATP when measuring flow.
The following instructions will help guide you through the setup screens of the
Certifier FA Plus and select the appropriate parameters and save a
configuration for the Esprit Ventilator.
Measurement Selection Screen
Use this screen to add or remove parameters from the Certifier FA Plus screen.
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G