Oxygen Flow Sensor

REF 580-1000-02 G
Figure 9-68: Filter Heater Tube Connection
Follow these steps to remove the oxygen flow sensor (Figure 9-69):
1. Remove the top enclosure (see "Top Enclosure" on page 9-4).
2. Remove the enclosure brace.
3. Completely remove the 4-mm screws and clamp from the oxygen flow
sensor.
4. Cut the tie wrap holding cable O
disconnect the cable.
5. Gently push the oxygen flow sensor and the aluminum sleeve toward
each other, then tilt the flow sensor up at the oxygen valve and lift out.
Esprit® Ventilator Service Manual © Respironics, Inc.
Component Removal/Installation
F/S to the oxygen flow sensor, then
2
Chapter 9
Filter
heater
assembly
Remove
silicone
tube from
tube nipple
9-57