1-4
Unit of
Measurement
Pressure
-25 to 150 cmH
100 psi
Flow (standard)
0 to 300 SLPM
Torque
5 to 25 in-lbs.
Volume (STP)
0 to 10 L STP
Voltage
DC: ± 5 to 50 V
AC: 2 to 300 V
Table 1-2: Test Equipment Specifications
Esprit® Ventilator Service Manual © Respironics, Inc.
Range
O 0 to
± 1% of reading or ± 0.20 cmH
2
is greater). ± 2% of reading @ -1 to 38ºC (30
to 100ºF)
± 2% of reading ± 0.20 SLPM (whichever is
greater)
± 1 in-lbs of reading
± 2% of reading ± 0.20 L STP
± 2% of reading
Accuracy
O (whichever
2
REF 580-1000-02 G