Oxygen Sensor (Test 7, Step 1 - Respironics ESPRIT Service Manual

Ventilator
Hide thumbs Also See for ESPRIT:
Table of Contents

Advertisement

Oxygen Sensor
(Test 7, Step 1)
REF 580-1000-02 G
This step establishes 100% oxygen at 200 LPM for one second, and 1 LPM for
two minutes, to calibrate the oxygen sensor.
Diagnostic code 3113 indicates a bad oxygen sensor or oxygen sensor cable.
The ventilator must pass EST following oxygen sensor replacement.
Esprit® Ventilator Service Manual © Respironics, Inc.
Appendix A
Red = air flow
Blue = O
flow
2
Black = no flow
A-15