Chapter 8
Performance Verification
Returning Ventilator
to Operation
8-40
5. With AC still disconnected, cycle power to the ventilator and allow it to
come up in normal ventilation.
6. Verify that the ventilator operates in normal ventilation, the message
Backup Battery On is displayed in the Alert window, the Battery In
Use indicator is on, and the Battery Low indicator turns off.
7. Reconnect the AC power cord. Verify that the ventilator continues
operating uninterrupted, the Mains indicator turns on, and the Battery
In Use indicator turns off.
8. Verify that the Battery Charging indicator lights within 30 seconds.
9. The backup battery test is complete.
Follow these steps to return the ventilator to operation after a successful
performance verification:
1. Remove all test equipment, tools, and materials from ventilator.
2. Simultaneously press the ALARM RESET and 100% O
front panel for approximately five seconds while turning ventilator
power on.
3. Touch OK to enter diagnostic mode.
4. Touch User Config on the diagnostic screen.
5. Touch Compliance to enable (white background) the compliance
factor.
6. Touch Bkup Battery to enable (white background) this feature.
7. Correct the date, time, and altitude if necessary.
8. Rerun electrical safety test if any repairs were made.
9. Reconnect the original patient circuit and inspiratory bacteria filter to
the ventilator.
10. Run SST in Adult Mode.
11. Turn the alarm volume control knob fully clockwise.
12. Turn ventilator OFF.