Oxygen Regulator

1. Disconnect the bacteria filter, patient circuit tube, and oxygen sensor
tee from the ventilator inspiratory outlet.
2. Connect the oxygen source to the ventilator.
3. Use a pressure gauge capable of measuring 100 PSI, and set the
pressure range to 100 PSI. Zero the gauge.
4. Connect the gauge to the oxygen valve pressure port (Figure 5-12).
Oxygen valve
5. Touch Blower to de-energize (gray background) the blower.
Esprit® Ventilator Service Manual © Respironics, Inc.
Check that Analyzer and Oxygen
the Flow to:
Flow Displays Read:
100 LPM
120 LPM
165 LPM
148.5 to 181.5 LPM
0 LPM
Figure 5-12: Oxygen High Pressure Connection
90 to 110 LPM
108 to 132 LPM
0 ± 0.1 LPM