Summary of Contents for Respironics System One Heated Humidifier

  • Page 1 System One Heated Humidifier USER MANUAL...
  • Page 2 © 2009 Respironics Inc. and its affiliates. All rights reserved.
  • Page 3: Table Of Contents

    Home Cleaning: Water Tank ..........................9 Home Cleaning: Humidifier Base ........................9 Hospital and Institution Cleaning ........................10 Service ..................................10 Traveling with the System ..........................10 Troubleshooting ..............................11 Specifications ................................13 Disposal ..................................14 How to Contact Respironics ..........................14 EMC Information ..............................15 Limited Warranty ........................... Back Page...
  • Page 4: Intended Use

    Have any damaged parts replaced before continuing use. • Periodically inspect the humidifier power cord for signs of wear or damage. If it becomes worn or damaged, contact Respironics or your home care provider for a replacement.
  • Page 5: System Features And Contents

    The fill lines indicates the maximum water level for safe operation. (Found on front and both sides of tank) note: Read all instructions before using the humidifier. note: This humidifier is specified for use with Respironics REMstar and BiPAP therapy devices only.
  • Page 6: Connecting The Therapy Device

    Connecting the Therapy Device 1. When using a humidifier, the patient circuit attaches to the air outlet port on the humidifier. 2. To connect the therapy device to the humidifier, first you must remove the side cover on the therapy device. Press in the tab on the side cover and pull the cover away from the unit.
  • Page 7: Disconnecting The Therapy Device

    Disconnecting the Therapy Device CauTion: To avoid spilling, do not disconnect the humidifier from the therapy device with water in the tank. Remove the water tank from the humidifier before disconnecting the therapy device. 1. Grasp the release latch on the bottom of the humidifier. 2.
  • Page 8 3. Press the tab in the hole on top of the tank in toward the front of the tank. Gently remove the tank lid from the tank base and set aside. Empty any remaining water from the base of the tank. Hinges Tank Lid Middle Seal...
  • Page 9 5. Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base. Close the lid until the tab on the lid snaps back under the lip in the tank base. Inspect the tank. When it is closed correctly, the lid should be seated completely on the middle seal and it should sit snugly on the tank base so the tab can easily snap back in place.
  • Page 10 13. The Humidifier icon and current Humidifier number setting will light up on the device. Display Screen Humidifier Icon & Number Settings Control Wheel/Push Button Ramp Button note: If you are having trouble with your mask, refer to the instructions included with the mask. 14.
  • Page 11: Home Cleaning: Water Tank

    Home Cleaning: Water Tank WaRninG: Before cleaning the humidifier, always remove from the therapy device. 1. Turn the humidifier setting to 0, turn the therapy device off, and allow the heater plate and water to cool. WaRninG: Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank.
  • Page 12: Hospital And Institution Cleaning

    Following disinfection, visually inspect each piece. Discard and replace any damaged parts. Service The humidifier does not require routine servicing. If any part of the humidifier is worn or damaged, contact Respironics or your home care provider. See the Troubleshooting section later in the manual for additional information.
  • Page 13: Troubleshooting

    The tubing is not Remove your mask and tubing and check for kinks or tears. If it is connected correctly and torn or damaged, contact your home care provider or Respironics doesn’t seal properly. for replacement tubing and/or mask. If the tubing is not damaged, reattach your tubing, turn on the airflow, and check to make sure you do not still feel air coming out of the port area.
  • Page 14 Route the tubing behind the bed’s headboard. If the humidifier does not operate correctly after falling, contact your home care provider or Respironics. note: For information on troubleshooting your therapy device, see the manual included with your therapy device.
  • Page 15: Specifications

    IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment EN ISO 8185:2007 General Requirements for Humidification Systems Electrical (When the heated humidifier is used with a Respironics therapy device) AC Power Consumption: 100 – 240 VAC, 50/60 Hz, 2.1 A DC Power Consumption: 12 VDC, 5.0 A...
  • Page 16: Disposal

    Dispose of this device in accordance with local regulations. How to Contact Respironics To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the Customer Service department at 1-800-345-6443 or 1-724-387-4000. You can also use the following address: Respironics, Inc.
  • Page 17: Emc Information

    EMC information Because this heated humidifier cannot operate without using a Respironics therapy device, refer to the device’s User Manual for additional EMC information. ’ : This device is intended for use in the electromagnetic uiDance anD anufacturer eclaration lectromaGnetic miSSionS environment specified below.
  • Page 18 ’ : This device is intended for use in the electromagnetic uiDance anD anufacturer eclaration lectromaGnetic mmunity environment specified below. The user of this device should make sure it is used in such an environment. iec 60601 t mmunity ompliance lectromaGnetic nvironment uiDance...
  • Page 19: Limited Warranty

    Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of two (2) years from the date of sale by Respironics, Inc. to the dealer.
  • Page 20 1058288 Manufactured for: Respironics Inc. 1057395 1001 Murry Ridge Lane JR 5/29/2009 Murrysville, PA 15668 USA EN-DOM...