Configuration Menu; Saving Configurations - Respironics ESPRIT Service Manual

Ventilator
Hide thumbs Also See for ESPRIT:
Table of Contents

Advertisement

Chapter 8
Performance Verification
8-10
2. Select Flow Rate (refer to Figure 8-6) and verify Auto-Triggering is not
checked.
Figure 8-6: Trigger Options Screen
3. Use the up or down direction arrows to select 20/Lmin for Start Trigger
and 2/Lmin for End Trigger (refer to Figure 8-6).
4. When the appropriate parameters are selected, touch OK (refer to
Figure 8-6).

Configuration Menu

Use this screen to save configurations and switch between different
configurations.

Saving Configurations

1. Touch the "Configuration" tab (refer to Figure 8-3).
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G