REF 580-1000-02 G
Figure 9-20: MMI PCB Connectors (9.5-in. GUI)
7. Remove the four screws holding the MMI PCB to the mounting plate.
8. Pull the MMI PCB away from the plate just enough to remove the two
screws that hold the cable (MMI J14) to connector J14, then
disconnect the cable and remove it from the U-brace.
9. Disconnect the cable (MMI J2) from connector J2.
10. Disconnect ribbon cables from the connectors J6 and J13 (touch
frame), then pull the MMI PCB completely out of the top enclosure.
NOTE:
When reinstalling the ribbon cables to J6 and J11 (overlay LEDs), tug on
the cable to ensure that it is properly seated. When reinstalling the ribbon
cable to J12 and J13, ensure that no connector pins are visible, then
snap cover shut.
Esprit® Ventilator Service Manual © Respironics, Inc.
Component Removal/Installation
Chapter 9
9-17