F.5
Programming via the handheld programmer
To activate PROGRAM mode, point the handheld programmer at the display (from a
maximum distance of 600 mm (ft) and press the Mode key.
①
②
③
• In Navigation mode the rightmost digit of the menu number flashes and the PROGRAM
icon is not visible.
• In Edit Mode the PROGRAM icon appears and flashes: press a number key to enter
parameter data. For example, press 3 to move to the third item in the current menu.
SITRANS LR460
Operating Instructions, 10/2021, A5E42035355-AB
Display
Max. 600 mm (2 ft)
Handheld programmer
F.5 Programming via the handheld programmer
Local operation interface
169