Parameter assignment
8.7 Maintenance settings (MAINT SET) (6.)
Acknowledge (ACK STATUS) (6.2.8.)
Read only. Displays the status of the Maintenance Alerts that have been acknowledged.
Values
To view via SIMATIC PDM open the menu View - Device Status (PROFIBUS PA) or View – Status
(HART), and click on the Maintenance tab and check the Sensor Lifetime Status window.
Acknowledge warnings (ACK)
Allows you to acknowledge either a Maintenance Required or a Maintenance Demanded
alert.
To acknowledge an alert via SIMATIC PDM:
1. Open the menu View - Device Status (PROFIBUS PA) or View – Status (HART), and click on
the Maintenance tab.
2. In the Sensor Lifetime window click on Acknowledge Warnings.
Options (read only)
8.7.3
Service interval (SERVICE) (6.3.)
Allows for scheduling of service inspections. To access these parameters via SIMATIC PDM,
open the menu Device – Maintenance and click on the Service Interval tab.
Time lapsed from last service (LAST SVC) (6.3.1.)
Read only. Time elapsed since device was last serviced. Can be reset to zero via the handheld
programmer (after performing a service).
112
NONE
ACK M REQ
ACK M DEM
BOTH
None
Maintenance required acknowledged
Maintenance demanded acknowledged
Both maintenance required and demanded
acknowledged
IGNORE
ACK WARNING
Operating Instructions, 10/2021, A5E42035355-AB
SITRANS LR460