ntroduction

Chapter 1
mportant Note: Read the information in this user guide completely
before using the COD165 thermoreactor . The manufacturer does not
acknowledge any responsibility for improper use of the equipment .
mportant Note: The labels on to the COD165 thermoreactor warn users
of the dangers to which they are exposed to during use or maintenance .
The labels must not be removed from the COD165 thermoreactor .
Warning: The COD165 thermoreactor operates at high temperatures and
special care is required during use .

Safety Rules

The heating plate may reach temperatures of 165 °C during the heating
and cooling phases .
Only glass containers that can withstand a temperature greater than 165 °C
should be used with the COD165 thermoreactor .

Cleaning

Always unplug the COD165 thermoreactor before cleaning it . The heating
plate must be cool before the instrument is cleaned . Use a damp cloth with
nonflammable, non-abrasive detergents .
Personal Protection
The COD165 thermoreactor operates at high temperatures and is designed
for use with glass vials containing concentrated acid . Appropriate PPE
Equipment (PPE)
should be worn at all times during operation .

Repair and Maintenance

For repair and maintenance, contact Technical Support . Within the United
States call 1 .800 .225 .1480 and outside the United States call 978 .232 .6000
or fax 978 .232 .6031 . In Europe, the Middle East and Africa, contact your
local authorized dealer . For the most current contact information, visit
www .thermo .com/contactwater .