Thermo Scientific
Observe the safety instructions of the disinfectant's manufacturer and the hygiene
supervisor.
Incompatible cleaners
Some device components are made of plastic. Solvents can dissolve plastics.
Strong acids or caustic solutions can cause plastic to become brittle. For
cleaning plastic components and surfaces, do not use solvents that contain
hydrocarbons, solvents with an alcohol content of more than10% or strong acids
or caustic solutions!
Decontamination or cleaning agents
The "Thermo Scientific" company or their agent is to be consulted if there is any
doubt about the compatibility of decontamination or cleaning agents.
NOTE
Special cleaning instructions for the light module
When cleaning a light module, refer also to the section below with extended
cleaning instructions.

Moisture-Sensitive Components

Sensitive components
Do not spray cleaning agent onto the controllers of the device. When wiping the device
clean, always make sure that moisture does not enter into these components.

Cleaning Exterior Surfaces

1. Thoroughly remove dirt residues and deposits using a solution of tepid water and
dishwashing agent.
2. Wipe the surfaces clean using a clean cloth and clear water.
3. Then, wipe the surfaces dry using a clean cloth.
Cleaning and Disinfection |