11.Technical specifications

Operating
10 - 40 °C / 50 - 104 °F
conditions:
15-90 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15-90 % relative maximum humidity
Weight:
251 g (including batteries)
Dimensions:
124 x 92 x 61 mm
Cuff size:
from 22 - 42 cm according to the cuff
sizes (see «Selecting the correct cuff»)
Measuring proce-
oscillometric, corresponding to Korotkoff
dure:
method: Phase I systolic, Phase V
diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
0 - 299 mmHg
range:
Resolution:
1 mmHg
Static accuracy:
within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
 4 x 1.5 V alkaline batteries; size
 Mains adapter DC 6V, 600 mA
Battery lifetime:
approx. 400 measurements
(using new batteries)
IP Class:
IP 20
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000 measure-
ments, whichever comes first
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.